Comparability of automated drusen volume measurements in age-related macular degeneration: a MACUSTAR study report

Abstract

90% in the three modalities. In eAMD, we observed highest sensitivity in the denser Spectralis scan (68.1). The two different Spectralis modalities showed a significantly higher agreement in quantifying drusen volume in iAMD (ICC 0.993 [0.991-0.994]) than the dense Spectralis with Cirrus scan (ICC 0.807 [0.757-0.847]). Formulae for drusen volume conversion in iAMD between the two devices are provided. Automated drusen volume measures are not interchangeable between devices and softwares and need to be interpreted with the used imaging devices and software in mind. Accounting for systematic difference between methods increases comparability and conversion formulae are provided. Less dense scans did not affect drusen volume measurements in iAMD but decreased sensitivity for medium drusen in eAMD.Trial registration: ClinicalTrials.gov NCT03349801. Registered on 22 November 2017.

Publication
Sci Rep